Documentation Index
Fetch the complete documentation index at: https://docs.mykit.info/llms.txt
Use this file to discover all available pages before exploring further.
Why calibration matters
The watch detects shots using its accelerometer. Every firearm has a different recoil signature — a .22 LR has far less impulse than a .308. Without calibration, the watch may under-count shots (threshold too high) or trigger on movement that isn’t a shot (threshold too low or debounce too short). Two parameters control detection:| Parameter | What it does | Default |
|---|---|---|
| Threshold (g) | Minimum acceleration spike to count as a shot. Lower = more sensitive. | 2.5 g |
| Debounce (ms) | Minimum time between counted shots. Prevents double-counting from recoil resonance. | 150 ms |
Open calibration settings
- Go to Armory and click on any firearm to open its detail page.
- Scroll to the Watch Calibration section.
Manual calibration
Use manual mode when you already know good values from prior experience.- Select Manual in the calibration section.
- Enter a Threshold value (g above baseline).
- Enter a Debounce value (ms).
- Click Save.
Watch-assisted calibration
Use watch-assisted mode to derive the debounce value automatically from a real firing session.- Select Watch in the calibration section.
- Choose a watch from the dropdown and click Queue Calibration Drill.
- Run the drill on the watch — fire all rounds.
- Return to the calibration section and click Derive from Last Session.
- The debounce field auto-fills based on 80% of the minimum inter-shot interval observed.
- Adjust the Threshold manually if shots are being missed or double-counted.
- Click Save.
Threshold cannot be derived from timestamps alone — it depends on the accelerometer signal amplitude, which only the watch records. Start with 2.5 g and adjust up if you see phantom shots, or down if shots are being missed.
Calibration and the Send modal
The Send to Watch modal shows a warning if you select a firearm with no calibration data. This is not a blocker — the watch will use defaults — but accuracy will be lower than a calibrated firearm.Recommended workflow
- Fire 10–15 rounds at a normal pace with a new firearm.
- Use Derive from Last Session to set debounce.
- Check the split times in the Performance tab — they should match your actual cadence.
- If shots are missing, lower the threshold by 0.5 g and repeat.
- If you see extra phantom shots, raise the threshold by 0.5 g.

